种植体中央螺丝扭矩衰减的回顾性临床研究

虞颖娟1,朱文卿2()   

  1. 1 常州市口腔医院口腔种植科,江苏常州(213003)
    2 南京医科大学附属口腔医院口腔特诊科,南京医科大学口腔疾病研究江苏省重点实验室,江苏南京(210029)
  • 收稿日期:2023-08-29 出版日期:2024-01-28 发布日期:2024-01-22
  • 通讯作者: 朱文卿 Tel:(025)69593453 E-mail:zwqswxxx@163.com
  • 基金资助:
    国家自然科学基金青年项目(82201096);江苏省科教能力提升工程——江苏省研究型医院(YJXYYJSDW4);江苏省医学创新中心(CXZX202227)

A retrospective clinical study on the torque loss of implant abutment screw

YU Yingjuan1,ZHU Wenqing2()   

  1. Department of Oral Implantology, Changzhou Stomatological Hospital, Changzhou 213003, China
  • Received:2023-08-29 Online:2024-01-28 Published:2024-01-22

摘要:

目的 研究种植体中央螺丝扭矩衰减的临床影响因素,探讨预防中央螺丝松动的适宜加力时机。方法 收集2019年12月至2023年6月期间于常州市口腔医院种植科就诊的奥齿泰种植固定修复术后复查的患者。测量并记录中央螺丝扭矩衰减值以及可能的影响因素,包括性别,牙位,种植体直径,种植体负荷天数,共同支持的种植体数、牙冠数和负重比。对可能的影响因素进行皮尔森相关性及多元线性回归分析。结果 本研究共收集73例患者的125颗种植体。相关性分析及回归分析显示,种植体负荷天数对中央螺丝扭矩衰减值有显著影响(r=0.616,P<0.001),其余自变量与扭矩衰减值无显著相关性(P>0.05),据此建立的线性回归模型具有统计学意义(R2=0.475,P<0.001)。结论 奥齿泰标准型种植体在正常负荷下,种植体负荷天数可以显著正向预测中央螺丝扭矩衰减值。

关键词: 口腔种植, 中央螺丝, 扭矩衰减

Abstract:

Objective To study the clinical influencing factors of the torque loss value of implant abutment screw and explore theappropriate timing to prevent abutment screw loosening. Methods Patients who visited Implant Department of Changzhou Stomatological Hospital for examination after implantation from December 2019 to June 2023 were collected, and relevant clinical factors were measured and recorded, including patients’ gender, tooth position, implant diameter, the specification of abutment screw, the number of co-supported implants and bridges, the ratio of loading, loading time of the crown, torque loss value of abutment screw. The Person correlation analysis was performed to filter out independent variables that were highly correlated with the torque loss value of the central screw, and the linear regression analysis was performed. Results This study collected a total of 125 implants from 73 patients. According to Person correlation analysis and multiple linear regression analysis, implant crown load days had a significant impact on the central screw torque attenuation value(r=0.616, P<0.001), while the other independent variables had no significant correlation(P>0.05). The regression model was statistically significant(R2=0.475, P<0.001). Conclusion The loading days of Osstem standard implant can significantly predict the torque loss value of the abutment screw under normal load.

Key words: oral implant, abutment screw, torque loss